XG1.D Series

Motorize Door Pre-vacuum Sterilizer

Features

The sealing door adopts the technology of electrical lifting and sealing with compressed air to reduce labor cost.
 
Touch screen as interface makes the operation more easy.
 
The parameters such as time, temperature, pressure, etc in process can be displayed and adjusted by screen, these parameters also can be printed by built-in printer.
 
Program is fully automatic after starting, manual operation is also available.
 
It adopts SIEMENS PLC to control the process, high efficiency with the function of reliability and fl exibility.
 
The vacuum step is completed by pulse vacuum with feeding steam several times so as to ensure air exhaustion is over 99%, to avoid temperature dead angle and small volume effect to ensure the good result of sterilization.
 
Critical control components and valves are high quality imported to improve the reliability and stability of machine.
 
Shelf or loading trolley and transfer trolley is standard accessories.
 
Air compressor is optional.
 
Material for chamber, jacket and piping can be customized.

Safety features

The process can not be started when the door is unlocked.
 
The door can not be opened if there is pressure in chamber for ensuring the safety of both operators and machine.
 
For double door type, only one door can be opened at same time.
 
The filtering precision of air filter can reach 0.2 micron, to ensure no second pollution after sterilization.
 
Password is needed for parameters modification.
 
Emergency switch can stop any abnormal operation occurs.
 
Safety valves will release pressure of chamber and jacket if pressure is out of range.
 
Various alarms can stop operation and indicate the trouble spot.

Technical parameters

 

Rated working pressure: 0.21Mpa
Rated working temperature: 132℃
Range of pulse times: 0~99 times
Range of time for sterilization: 0~9999 seconds
Range of time for drying: 0~9999 seconds
Safety valve: open pressure: 0.23Mpa(jacket)
close pressure: 0.21Mpa
Safety valve: open pressure: 0.23Mpa(chamber)
close pressure: 0.21Mpa
Pulse vacuum range: positive pressure 0.065Mpa~0.09Mpa
negative pressure –0.05Mpa~-0.09Mpa
Air exhausted: >99% (≥3 times of pulse vacuum)