F96S
Fluorescence Spectrophotometer
Performance Index of the Instrument (Specifications)
Excitation light source | High brightness LED |
Excitation filter | Interfere Filter, bandwidth 10nm |
Emission monochromator |
C-T configuration diffraction grating monochromator Emission wavelength range (EM): 200nm ~ 650nm, Bandwidth: 10nm |
(Extend the monochromator to Em200-900 is optional.) | |
Emission wavelength accuracy: ± 1nm | |
Emission wavelength reproducibility: ≤ 0.5nm | |
S∕N ratio | S/N≥90(Using 1cm quartz sample cell, measure the signal |
noise ratio of Raman spectrum of water ) | |
Detection limit | 1 × 10-10 g / ml quinine sulfate solution |
Linearity (γ) | ≥0.995 |
Repetitive peak intensity | ≤1.5% |
Zero drift | ≤0.3(within 10min) |
Upper limit change of indicating value | ≤1.5% (Within 10 minutes) (displaying value≥50) |
Power type | 220V±22V; 110V±22V |
Dimensions | 442×392×250(mm) |
Weight | Net weight 10kg |
Gross weight 12kg |