MicroPette Plus
Autoclavable Pipette (Adjustable & Fixed)
Fully autoclavable
The MicroPette Plus pipettes are fully Autoclavable enable easy cleaning to reduce the risk of contamination. Steam autoclaving
can be performed at 121 °C, 1 bar for 20 minutes. After autoclaving the pipette must be cooled down and left to dry for
12 hours before use.
It is recommended to check the performance of the pipette after each autoclaving. Greasing and sealing of the pipette piston after every 10th autoclaving will secure an enhanced function.
Features
MicroPette plus
• Fully autoclavable
• Ergonomic design provides excel lent operating experience
• Easy-to-read volume display
• The pipettes cover the volume range from 0.1μL to 10mL
• Easy calibration and maintenance
• Manufactured from innovative materials
• Each MicroPette Plus supplied with an individual calibration certificate according to ISO8655.
MicroPette plus multi-channel
• 8 and 12 channel pipettes are appropriate for 96-well plates
• Dispensing head rotates for effortless pipetting convenience
• Individual piston and tip cone assemblies allows for easy repair and maintenance
• Compound material-made tip cone secures high sealing performance
• Compatible with most universal tip brands
Specifications
Single-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.1-2.5μL | 0.05μL | 2.5μL | 2.50% | 0.0625 | 2.00% | 0.05 | |||||
1.25μL | 3.00% | 0.0375 | 3.00% | 0.0375 | |||||||
0.25μL | 12.00% | 0.03 | 6.00% | 0.015 | |||||||
0.5-10μL | 0.1μL | 10μL | 1.00% | 0.1 | 0.80% | 0.08 | |||||
5μL | 1.50% | 0.075 | 1.50% | 0.075 | |||||||
1μL | 2.50% | 0.025 | 1.50% | 0.015 | |||||||
2-20μL | 0.5μL | 20μL | 0.90% | 0.18 | 0.40% | 0.08 | |||||
10μL | 1.20% | 0.12 | 1.00% | 0.1 | |||||||
2μL | 3.00% | 0.06 | 2.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 0.60% | 0.3 | 0.30% | 0.15 | |||||
25μL | 0.90% | 0.225 | 0.60% | 0.15 | |||||||
5μL | 2.00% | 0.1 | 2.00% | 0.1 | |||||||
10-100μL | 1μL | 100μL | 0.80% | 0.8 | 0.15% | 0.15 | |||||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||||
10μL | 3.00% | 0.3 | 1.50% | 0.15 | |||||||
20-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||||
20μL | 3.00% | 0.6 | 1.00% | 0.2 | |||||||
50-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||||
100-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||||
100μL | 2.00% | 2 | 0.70% | 0.7 | |||||||
200-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||||
200μL | 0.90% | 1.8 | 0.30% | 0.6 | |||||||
1000-5000μL | 50μL | 5000μL | 0.50% | 25 | 0.15% | 7.5 | |||||
2500μL | 0.60% | 15 | 0.30% | 7.5 | |||||||
1000μL | 0.70% | 7 | 0.30% | 3 | |||||||
2-10mL | 0.1mL | 10mL | 0.60% | 60 | 0.20% | 20 | |||||
5mL | 1.20% | 60 | 0.30% | 15 | |||||||
2mL | 3.00% | 60 | 0.60% | 12 | |||||||
8-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
12-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
Fixed Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
5μL | - | 5μL | 1.30% | 0.065 | 1.20% | 0.06 | |||||
10μL | - | 10μL | 0.80% | 0.08 | 0.80% | 0.08 | |||||
20μL | - | 20μL | 0.60% | 0.12 | 0.50% | 0.1 | |||||
25μL | - | 25μL | 0.50% | 0.125 | 0.30% | 0.075 | |||||
50μL | - | 50μL | 0.50% | 0.25 | 0.30% | 0.15 | |||||
100μL | - | 100μL | 0.50% | 0.5 | 0.30% | 0.3 | |||||
200μL | - | 200μL | 0.40% | 0.8 | 0.20% | 0.4 | |||||
250μL | -- | 250μL | 0.40% | 1 | 0.20% | 0.5 | |||||
500μL | - | 500μL | 0.30% | 1.5 | 0.20% | 1 | |||||
1000μL | - | 1000μL | 0.30% | 3 | 0.20% | 2 | |||||
2000μL | - | 2000μL | 0.30% | 6 | 0.15% | 3 | |||||
5000μL | - | 5000μL | 0.30% | 15 | 0.15% | 7.5 |