产品列表

所有产品 (A~Z)

联系我们

 

 销售热线:

 0571-87977725

 0571-87977729

 

 电子邮箱:

 hzckmy@gmail.com

 

 

 
您的位置: 首页 > 产品 > 专用仪器 > 公路检测仪器 > 其它 > PAV-1产品

PAV-1

Asphalt Pressure Aging Vessel System

Summary


This system is designed and made as per industrial standards T 0630-2011 Accelerated Aging Test for Bitumen by Pressure Aging Vessel in JTG E20-2011 Standard Test Methods of Bitumen and Bituminous Mixtures for Highway Engineering and SH/T 0774-2005 Standard practice for accelerated aging of asphalt binder using a pressurized aging vessel (PAV). It also refers to standards ASTM D6521 and AASHTO PPI. It covers the accelerated aging (oxidation) of asphalt binders by means of pressurized air and elevated temperature.


 
Main technical features


1. This system is composed of PAV-1-1 pressuring aging vessel and PAV-1-2 aging vacuum vessel.
2. This system is equipped with a PC. The high pressure aging test is controlled by a PC. All test procedures can be controlled by a PC.


 
Main technical specifications

 

(1) PAV-1-1 pressuring aging vessel


1.Electricity
Power supply:   AC(220±10%)V、50Hz
Rated power:     1100W
Rated current:    10A
2.Vessel
Volume:    13L
Medium:    Air
Rated working temp.:     110℃±5℃
Rated working pressure:   2.1MPa
Designed pressure:     2.2 MPa
Water pressure:     2.75 MPa                                                 
Safety valve opening pressure:   2.15 MPa
Explosive pressure for rupture disk:  2.6 MPa
3. Pressure controller
Model:     WP962-2(WP-D90)
Precision:   ±0.01MPa
4. Temperature and time controller     Model WP962-1(WP-D90)
Temperature control precision:    ±0.1℃
Grading:   64×1080min
Outage maintaining function:     Yes
5.Dimension :    630mm×450mm×500mm


(2)PAV-1-2 aging vacuum vessel
 

1.Electricity
Power supply:    AC(220±10%)V、50Hz
Rated power:      900W
Rated current:     10A
2.Vessel
Volume:    13L
Medium:    Air
Rated working temp.:   180℃±5℃
Rated working pressure:    Absolute pressure 15kPa
Pressure control mode:     Electric contact pressure gauge
3. Temperature and time controller:  Model XMT-E55T2
Temperature control precision:    ±0.1℃
Time control mode:   Button
4. Dimension:  560mm×390mm×350mm