GC126N

Gas Chromatograph

Technical Features

● The host machine adopts a 7-inch color touch screen with friendly man-machine interface.

● Computer backcontrol and host touch screen to achieve synchronous bidirectional control.

● The multi-core, 32-bit embedded hardware system ensures the reliable operation of the instrument.

● Carrier/hydrogen/air flow (pressure) digital display.

● Gas shortage alarm protection function; Heating control protection function (when opening the door of the column box, the motor of the column box fan and the heating system will shut down automatically).

● Split flow/split ratio can be automatically controlled to save carrier gas.

● Configure automatic sampler installation and positioning interface to match automatic sampler of various specifications.

● Data acquisition is a standard dual-channel data acquisition card with a sampling time of 50ms.

● Using logarithmic amplification plate, detection signal no cut-off value, synchronous external trigger function, can be started by external signals (automatic sampler, thermal analyzer, etc.) at the same time the host and workstation.

● It has perfect system self-check function and fault automatic identification function.

● With eight external event extension function interface, can be selected with a variety of control valves, and according to their own set time sequence work.

● The external link mode is network port connection (RJ45), which provides convenience for remote control of data.

 

Column temperature box:

● Temperature range: 5℃ ~ 400℃ at room temperature

● Temperature control accuracy: ±0.1℃

● Program heating: stage 9/10 platform

● Total program time: 9999.9 min

● Maximum heating rate: 60℃ / min

 

Sampler:

● Temperature range: 7℃ ~ 420℃ at room temperature

● Fill column, split/split sampler can be installed

● Working under constant pressure mode

● No more than three independent sampling systems are connected simultaneously

 

Detector:

● A maximum of 2 units are installed simultaneously. FID, TCD (optional)

● Flow/pressure is displayed directly on the screen.

● Temperature setting: Max. 420℃

● Hydrogen flame ionization detector (FID)

High voltage switch control

Baseline signal display

Ignition coil control

Detection limit: ≤ 3×10-12 g/s

Sample: N-hexadecane (minimum detection quantity: 3pg/s)

Dynamic range: 107

 

Heating area:

● In addition to the column box independent heating area, there are six heating areas. 2 injector heating zones, 2 detector heating zones and 2 auxiliary heating zones.

● The maximum operating temperature of the auxiliary heating area can reach 400℃

● Supply voltage: AC220V±22V, 50Hz±0.5Hz
● Power: 3000W

 

Optional

● Thermal conductivity detector (TCD) :

Bridge voltage switch control

Bridge current setting: 0 ~ 220 mA

Sensitivity: 5000 mV•mL/mg

Dynamic range: 105

● Packed column

● Capillary column