CI-B5

BOD METER

 

Applications

This instrument is to test BOD value in surface water, sewage, industrial wastewater especially used to measure waste loadings at wastewater treatment plants and to examine the efficiency of wastewater treatment.

 

Principle

Designed by the respirometric method, CI-B5 BOD Tester simulates the degradation process of organics in the nature. Viz. the air in the bottle above the sample contains 21% oxygen and replenishes the dissolved oxygen used by the bacteria. Carbon dioxide is made when microorganisms oxidize organic matter in the sample. The carbon dioxide must be removed by the potassium hydroxide from the system so it does not interfere with the measurement. Pressure sensors monitor air pressure in the sample bottles. The pressure drop correlates directly to BOD and BOD is correspondence with the quantity of oxygen consumed by bacteria. The instrument will deal with this correlation and then the BOD value will be displayed on the screen.

Technical Specifications

1.  Accuracy: ±5% (full range)

2. Sample range: liquid sample, including dissoluble organic sample, e.g.   released sewage and river water

3. Measurement range: 0 to 35, 0 to 70, 0 to 350, 0 to 700mg/L (no need to dilute)

4. 700 to 4000 mg/L (need to dilute)

5. Sample quantity: ≤6

6. Volume: 500 ml

7. Repeatability: ≤5%

8. Resolution ratio: 0.1mg/L (0 to 35, 0 to 70mg/L)

                                   1mg/L (0 to 350, 0 to 700mg/L)

9. Display: LCD

10. Power source: AC 220, 50Hz

11. Dimensions: 35×40×35cm

 

Key Features

1. Conforms to GB7488-87

2. Display the trend curve

3. Printer output

4. Link to the compute through USB interface

5. Mercury-free devices to ensure the accuracy and safety of the BOD measurement

6. Real-time clock display

7. Automatic collection and storage of data

8. Depict BOD curves for the user to view

9. Link to the computer to print curves

10. No parameter and data loss after suddenly electricity cut-off

11. LCD humanization interface display, easy to operation

12. 1-30days of the test time, setting by the keys

13. Store data and update the result, observe the change rate of BOD in the LCD screen during any testing time, store 240 data points and 20 groups data of each sample.