BJ-2
Disintegration Tester
BJ-Ⅱ Disintegration Time Limit Tester
Disintegration time limit tester is a drug test instrument to test the disintegration time limit of tablets, capsules and pills in accordance with Chinese Pharmacopoeia.
Technical indexes:
Temperature prefabrication range: room temperature:~50℃, display resolution:0.1℃.
Temperature control accuracy: ±0.3℃
Timing prefabrication range: 1min~900min
Time control accuracy: ±0.5min
Work noise: <60db
Number of lifting baskets: 2 sets
Lifting frequency of basket: (30~31) times/min.
Lifting distance of basket: (55±1) mm
Minimum spacing from screen to beaker bottom: 25mm±2mm
Mesh size: standard configuration 2mm, (optional for 0.425mm and 1mm bore diameters)
Power supply:220V50Hz
Total power:600W
Overall dimensions:L*W*H 400mm*320mm*440mm
Main features:
◎Double-basket working position system measures disintegration time limit.
◎2 sets of baskets, which can conduct independent disintegration experiment respectively.
◎Electronic temperature sensor can display and monitor various points in water bath case and the temperature in the beaker. High-precision digital electronic sensor is adopted to ensure the high precision and high accuracy of water bath temperature without the need for calibration.
◎The instrument controls the bath temperature at 37.0℃ automatically (prescribed by Pharmacopeia). It can also reset the preset temperature again at any time.
◎The instrument sets the lifting time of baskets to be 15min automatically, which can also be reset at any time.
◎It adopts the computer control technology which is centered on SCM to control two performance parameters including bath temperature and working time intelligently.
◎The instrument can monitor bath temperature, give over-temperature alarm and protect itself automatically.