Product List

ALL PRODUCTS (A~Z)

Inquiry Now

  Email:

  hzck@com.hz.cn

  sales@cktrade.com.cn

 

 

   Fax:

  +86-571-87977703

 

 

  Tel:

  +86-571-87977725

  +86-571-87977729

  +86 18969038315

 

 

   Skype:

  ckskyphone

 

 

 

You are here: Home > Products > General > Autoclave > Low Temperature > PS-100Products

PS-100

Low Temperature Plasma Sterilizer

Features

PS-100 low temperature plasma sterilizers are for the sterilization of thermolabile materials and instruments. The advantages include cycle times that are less than one hour, no aeration time, no harm to people or the environment, protect valuable instruments etc.. So it is a completed safe sterilization method. It is widely used for medical operating
room, CSSD and other treatment-room.

Inapplicable materials and instruments

Stainless steel tube: inside diameter less than Φ1, length longer than 400mm
 
Moisture absorption (woody apparatus、fibrin、cotton fabric、gauze etc.)
 
Disposable instrument (disposable waterproof weave, disposable operativegown etc.)
 
Liquid or powder Incomplete articles
 
Dead end lumen
 
Implant
 
Can not sustain vacuum and the mark can only use am sterilizing equipments.

Main Features

Hydrogen peroxide injection system with high precision measuring pump, it can be adjusted the injection level, the reliability and the precision of injection can be ncreased. One time injection can be used for 50 cycles.
 
Hydrogen peroxide refrigerating system can realize low temperature storage. Safety and the stability of the chemical properties can be ensured.
 
High capacity of storage to store the initial data.
 
Colorful touch screen makes operation easier. All the parameters can be displayed and printed. Each step and any error can be displayed on touch screen.
 
Easy installation, only a 220V 50Hz power is needed.
 
All the necessary accessories and consumables can be offered , more convenience for customer.
 
Inapplicable materials and instruments
 

Technical parameters

 

Chamber Dimensions:  750mm x 450mm x 400mm(LxWxH) Volume: 135L
Overall Dimensions:  750mm x 450mm x 400mm(LxWxH) Volume: 135L
Weight:  400Kg
Microbicide: 60% hydrogen peroxide
Consumption:  4.4ml per standard cycle
Sterilization Time:  standard cycle takes 50 minutes
Sterilization Temperature:  50℃+/-5℃;
Chamber vacuum compression:  10~60pa
Power supply:  single-phase 220V 50Hz 
Power: 3Kw
Working condition:  10~40℃
Standard of HF Power source:  Frequency13.56MHz, Max power: 500W;
Installation space: Min space:  1200 ×1500 mm

Sterilization process

 

 
CODE PHASE(PH) VACUUM LEVEL TIME
1 VACUUM PH ≤66Pa(0 . 5 Torr) 13~15minutes
2 DIFFUSION PHI ≥1332Pa(10 Torr) 6minute
3 VACUUM PHII ≤66Pa(0 . 5 Torr) 6~8minute
4 PLASMA PH I ≤66Pa(0 . 5 Torr) 2 minute
5 DIFFUSION PHII ≥1332Pa(10 Torr) 6 minute
6 VACUUM PH ≤66Pa(0 . 5 Torr) 6~8 minute
7 PLASMA PH II ≤66Pa(0 . 5 Torr) 2 minute
8 HAUL-BACK PH 100kPa (760 Torr) 1 minute