YD300
Portable Water Hardness Meter
Introduction
●Portable meter for measurement of water hardness (the total concentration of calcium ion and magnesium ion).
● Electrode adopts model 601-F water hardness combination electrode with patent structure. It consists of measurement electrode, reference electrode and temperature probe. Replaceable measuring head, steady potential, fast response, easy to use.
●Meter adopts intelligent chip design, automatic calibration, automatic temperature compensation, time display, data storage, low voltage display, etc.
● Ability to switch among eight units: mmol/L, mg/L (CaCO3), mg/L (CaO), mmol/L (Boiler), mg/L (Ca). Three options of three units : ºfH(France)、ºdH (Germany)and ºeH(UK)from WH-Link communication software.
● Equipped with B1、B2 and B3 calibration solution.
● IP57 waterproof rating, equipped with portable case, calibration solution and other accessories.
Specification
Water Hardness | Measuring range | (0 ~ 10)mmol/L; (0 ~ 401)mg/L(Ca); |
(0 ~ 1000)mg/L(CaCO3); (0 ~ 100)ºfH(France Degree); | ||
(0 ~ 561)mg/L(CaO); (0 ~ 56)ºdH(German Degree); | ||
(0 ~ 20)mmol/L(Boiler); (0 ~ 70)ºeH (England Degree)。 | ||
Resolution | 0.01 and 0.1 water hardness units | |
Accuracy | ±5% FS | |
Temp. compensation | (5 ~ 50)℃((automatic)) | |
Calibration Solution | B1 Calibration Solution — 2.00×10-2 mmol/L | |
B2 Calibration Solution — 2.00×10-1 mmol/L | ||
B3 Calibration Solution — 2.00 mmol/L | ||
(B3Calibration Solution also can be used for the soaking solution of electrode) | ||
Calibration Mode | (a)B1/B2 Calibration—using B1 and B2 Calibration Solution,suitable for < 2.00×10-2 mmol/L low concentration water quality,for example, boiler | |
water. | ||
(b)Calibration—using B2 and B3 Calibration Solution,suitable forgeneral water quality. | ||
Temperature | Measuring range | 0 ~ 60℃ |
Resolution | 0.1℃ | |
Accuracy | ±0.5℃ | |
Other Technical Parameter | Data storage | 99 groups |
Content Storage | serial number of measuring value, measuring value, units, temperature, ATC or MTC, date, time and calibration way. | |
Two AA cells(1.5V×2) | ||
Power | Meter:(65×120×31)mm/180g; Carry Case: (255×210×50)mm/790g | |
Dimension and weight | IISO9001:2000 and CE | |
Quality & safecertificate | ||
working conditions | Ambient temperature | (5~35)℃ |
ambient humidity | ≤85%RH |
1. Model YD300 water hardness meter | 1set |
2.Model 601-S water hardness electrode( With two measruing heads and one screw plug ) | 1pc |
3. AA cell (1.5V ) | 2pcs (included) |
4.100mL hardness calibration solution ( B1, B2 and B3 ) | 1btl/each |
5.20mL test cup | 6pcs |
6. Test cup pedestal | 1pc |
7.Carring case | 1pc |
8. Operation manual | 1 set |
Accessory Category | Description | Specification |
Water hardness electrode | 601-S water hardness electrode | Suited for measurement of total concentration of calcium ion and magnesium ion, match YD300 meter, 8 pin connector |
Water hardness calibration solution | Water hardness calibration solution B1 | 2.00×10-2 mmol/L,500mL/250mL/100mL(bottle) |
Water hardness calibration solution B2 | 2.00×10-1 mmol/L,500mL/250mL/100mL(bottle) | |
Water hardness calibration solution B3 | 2.00 mmol/L,500mL/250mL/100mL(bottle) | |
Plasic test cup | Specification:20mL | |
Accessories | Test cup base |
Place test cup on the base |