Product List

ALL PRODUCTS (A~Z)

Inquiry Now

  Email:

  hzck@com.hz.cn

  sales@cktrade.com.cn

 

 

   Fax:

  +86-571-87977703

 

 

  Tel:

  +86-571-87977725

  +86-571-87977729

  +86 18969038315

 

 

   Skype:

  ckskyphone

 

 

 

You are here: Home > Products > Physical > Spectrophotometer > Double Beam Spectrophotometer > G-9 Series Products

G-9 Series

Double Beam UV-Vis Spectrophotometer

G-9 series double beam scanning UV-visible spectrophotometer is designed by Philes for professional customers who need high requirements of accurate measurement in the organic chemistry, biochemistry, pharmaceutical analysis, food testing, medicine and health, environmental protection, life sciences and other fields of scientific research and production. Using the latest technology and electronic control systems, optimizing the traditional optical design and structural design to make sure the accuracy, stability, durability while expand new functionality; original Heraeus deuterium lamp imported from German to ensure the stability and reliability.


Function:


1.Variable bandwidth, suitable for different samples and different test requirements
2.Photometric measurements: sample can be easily measured absorbance and transmittance at the specified wavelength. Multi-wavelength pinpoint
3.test can be simultaneously test up to 30 wavelengths
4.Quantitative analysis: automatically build standard curve, optional one-class and zero-class linear regression, second-class and third-class curve fitting, single and double wavelength calibration and three-point method; with store standard curve to facilitate future testing
5.Qualitative Analysis: The scanning speed is adjustable, filter position can be fine-tuned, spectrum processing including scaling, smoothing, filtering, peak detection etc
6.Dynamic testing, enzyme kinetics reaction rate.
7.DNA \ RNA, protein test function: built-in specified formula to get the final result directly
8.USB data interface with online dedicated PC workstation, with more powerful function and data processing capabilities.

 

Performance feature:


1.Light source correction mechanism, the light source will be locating the best position when self-testing each time so that it can fix the offset focus position possibly caused by vibration during the transportation and the user don’t have to debugging light source after change it
2.With rigorous optical system design and circuit control system design, strict selection of components, with high standards of accuracy, repeatability, low noise and low stray light indicators, and excellent the linear range and stability index
3.With open sample chamber, optional reflective sample holder, automatic 4-positon cell holder, 7-position and 16-position cell holder, solid sample holder, water bath and auto sampler etc suitable for different applications.

Specifications: 

 

Optical System Double beam, Czerny-Turner, Grating 1200line/mm
Wavelength Accuracy ±0.1 nm(656.1 nm D2),±0.3 nm (the whole wavelength range)
Wavelength Repeatability ±0.1nm
Wavelength Resolution 0.1nm
Photometric Range ­-4 to 4A ;  0-200%T;  -9999 to 9999C
Photometric Accuracy  ±0.002A(0~0.5A)
±0.004A(0.5~1A) 
±0.008A(1~2A)
±0.3%T(0~100%T)
Photometric Repeatability ±0.001A(0~0.5A)
±0.002A(0.5~1A) 
±0.004A(1~2A)
±0.15%T(0~100%T)
Stray light ≤ 0.02%T (220 nm,NaI;340 nm,NaNO2)
Baseline Flatness ±0.0008A(190~1100nm)
Stability  ±0.0004A/hr @500nm ,0A
Photometric noise ±0.0003A
Derivative Resolution >0.5
Data Output 320×240 array graphic LCD
  USB connect to upper software
Size 650×450×220
Weight 25kg