JS-780
Automatic Gel Imaging Analysis System
JS-780 Automatic Gel Imaging Analysis System |
Name & Specification |
JS-780 Automatic Gel Imaging Device |
SONY ICXCCD professional digital CCD |
Computer M6Z1212M 2/3 inches, large diameter, high permeability lens, F=1:1.2, 12.5-75mm |
Coating filter specially designed for various fluorescent dye of gel imaging feature , 1 bit light filter wheel(optional of manual 5 bits light filter wheel and automatic 5 bits light filter wheel controlled by computer) |
Gel cutting devices |
Gel Image Shooting Software |
Gel Imaging Analysis Software |
Technical parameters
JS-780 Automatic Gel Imaging Analysis System | |
Digital CCD | SONY ICXCCD professional digital CCD |
Detection Sensitivity | The detection of Nucleic acid as Low as 0.02ng |
Effective Pixel | 1392 x 1040 |
Lens |
Japan automatic zoom lens with high permeability F=1:1.2, 12.5~75mm, gel size on request free to zoom |
Data bits | 12bit(0-4096 gray-scale) |
Dynamic range | >three orders of magnitude |
Integration of Pixel | 1 x 1, 2 x 2, 3 x 3, 4 x 4 |
Pixel size | 4.6μm x 4.6μm |
SNR | ≥ 60dB |
Shutter control | Electronic shutter |
Exposure time |
Long exposure time from 4ms to 60 minutes, collect and store all images successively, possess image correction function |
Interface | Standard C interface |
Cutting device |
special observation and cutting devices (visual angle and the object is 90 °) |
Color filter film |
Coating filter specially designed for various fluorescent dye of gel imaging feature ,Standard configuration of 590nm(or 535nm, 620nm, 460nm) EB/SYBR Green, BP |
Light filter location |
1 bit light filter wheel (optional of manual 5 bits light filter wheel and automatic 5 bits light filter wheel controlled by computer) |
Case control |
Via case panel to realize the automatic control - zoom, focus, aperture, transmission UV lamp and reflector lamp |
Computer control |
highly programmed (computer controlled black box / power / UV and white light switch / iris / focus / focal length) |
Timing shutdown |
Shutdown timer for 15 minutes, effectively extending the UV lamp life and UV glass |
The way of Opening door | pull-down to open the door and drawer-platform |
UV transmission loading board |
Transmission wavelength: 302nm(254nm, 365nm (is optional)), size: 20cm x 25cm(special specifications can be customized) |
White light reflection | Cooling light, voltage 12V |
White light transmission loading-board |
Cooling light, voltage 12V, chassis panel internal folding 20cm x 25cm (special specifications can be customized) |
Quality authentication |
Only one of gel imaging systems passes national authoritative measurement authentication instruction, and obtains third class of Shanghai scientific advance prize |
Shape size | L50CM X W37CM X H82CM |
Application |
can be used for DNA / RNA gels, protein gels, blot autoradiography film plastic card, the enzyme Standard plate, thin layer chromatography plate, imaging and analysis of dish |