HFPD-69

Filter Press, HTHP, 500 mL, Double Capped Cell for Cement Testing, Nitrogen Pressure

Characteristics:

The HTHP (High Temperature, High Pressure) Filter Press is designed for cement and testing drilling fluids under elevated temperatures and pressures. The unit simulates various downhole conditions and provides a reliable method for determining the effectiveness of the material being tested.
The High Pressure-High Temperature (HPHT) Filter Presses are an efficient means of evaluating the filtration properties of drilling muds at high temperatures and pressures. The cells have a filtering area of 3.5 in2 and can be operated at pressures up to 900 psi with the compact, easily attached CO2 pressuring unit provided.
The HTHP Filter Press is equipped with a 500 mL sample cell and the working temperatures above 500°F (260°C), making it ideal for laboratory use. Also included is a regulator manifold system for connecting to bottled nitrogen for higher pressures up to1, 800 PSI (12,400 kPa).

Parameter

Filter Press, HTHP, 500 mL, Double Capped Cell for Cement Testing

Part No.

Model

Specification

Remark

171-03

HFPD-69

Double Capped Cell for Cement Testing

Maximum Temperature : 500oF (260oC)

Maximum Pressure Cell : 1800 psig (12.4 MPa)

Maximum Backpressure : 750 psig (5.17 MPa)

Sample Cell Volume : 500 ml

Receiver Volume : 100 ml

Filtering Area : 3.5 in2 (22.58 cm2)

Pressure Connection : Nitrogen or Carbon Dioxide

Heating Capacity :800 Watts

Power Requirement : 115/230 VAC, Frequency 50/60 Hz

 

System Diagram