Dynamic HTHP Filter Press, Model HDF-01
Dynamic HTHP Filter Press
Features:
The industry’s only true dynamic filtration system for conducting filter cake formation and permeability analysis for drilling fluids optimization.
The Dynamic High Temperature High Pressure (HTHP) filter press measures filtration volume and cake building properties under varying down hole conditions. A motor driven shaft fitted with propellers turns at varying speeds inside a standard 500 ml HTHP cell. RPM settings from 1 to 1600 RPM impart laminar or turbulent flow to the fluid inside the cell. By varying the shaft length, the shear stress may be increased or decreased. The power is driven to the stirring shaft by a timing belt that is easily accessible for quick adjustment and removal.
Other features include a variable‐speed motor controlled through a SCR controller. RPM is indicated on a digital tachometer. The test procedure is exactly the same as that in the conventional HTHP filtration test. The only difference is the fluid circulates inside the cell while filtrate is being collected. Since the filter medium is conventional disk material, the results are fully comparable with those of other laboratories or for comparing historical trends.
Competitive Advantage:
The Dynamic HTHP Filter Press’s unique design is touted for providing you with the most representative data possible.
The ability to vary the propeller used, the length of the shaft and the RPM allow you to more accurately resemble the wellbore.
Because of it’s flexibility, Dynamic HTHP Filter Press is very useful for a variety of applications, including drilling fluids, drill‐in fluids, and completions.
In addition, Dynamic HTHP Filter Press is very economically priced compared to competitive models.
Features and Benefits:
‐ Maximum Pressure: 1,250 PSI (8.6 MPa) static; 800 PSI (5.5 MPa) dynamic
‐ Variable speed electric motor, ½ HP permanent magnet DC
‐ Auxiliary pipe connection for the cell top cap. Plug may be removed to add additional fluid additives.
‐ Safety calibrated rupture disk guards against overpressure
‐ Motor & stirring shaft RPM operate at a one‐to‐one ratio
‐ Adjustable thermostat from 100 to 500°F
‐ Optional filter permeabilities available
‐ Radial/axial flow propeller
‐ Adjustable propeller to change distance above filter media
Parameter
Dynamic HTHP Filter Press
|
Part No.
|
Model
|
Specification
|
Remark
|
172-50
|
HDF-01
|
The digital system is accurate and controllable.
Working pressure : 1800 psig
Maximum temperature : 500°F(260℃)
Maximum differential pressure :500+ psig
(actual maximum limited by core strength)
Maximum power requirements : 1500 watts
Heater power : 1200 watts
Power supply :120/240V, 50/60 Hz
Sample volume :500 cc
Filtrate volume :50 cc
Shear bob drive :¼ hp motor with belted magnetic drive;
no dynamic seals to wear out
Shear rate constant : 2693 1/s per rpm (no filter cake)
Initial shear rate range (with standard bob) :9, to 269 1/s
Dimensions : 12.5 x 17.75 x 28 in (305 x 455 x 711 mm)
Weight :170 pounds (77 kg)
|
|
System Diagram