Product List

ALL PRODUCTS (A~Z)

Inquiry Now

  Email:

  hzck@com.hz.cn

  sales@cktrade.com.cn

 

 

   Fax:

  +86-571-87977703

 

 

  Tel:

  +86-571-87977725

  +86-571-87977729

  +86 18969038315

 

 

   Skype:

  ckskyphone

 

 

 

You are here: Home > Products > Specialty > Agriculture&Food > HPR-1DProducts

HPR-1D

Hand-held Pesticide Residue Meter

Overview

This Handheld Pesticide Residue Test is portable, compact size and convenient to carry, 

adopts enzyme value method and shows the result of value. The pesticide residue is out of limits 

if 50% is positive, the higher of the value, the greater of the amount of residue.

 

Principle of Instrument

This Pesticide Residue Rapid Test is based on cholinesterase hydrolyzing acetylcholine. 

Its hydrolysate impacts DTMB, forming the yellow coupling things. Organophosphate pesticide 

can control enzyme. The changes of absorption value can show the degree of controlled enzyme 

and then reflect the pesticide residue in samples.

 

Area of Application

Detect products: agricultural and sideline products, daily food, seafood and its products and so on.

Detect object: rapid quantitative testing of organophosphate and carbamate pesticide residues.

Applicable units: departments of industry and commerce, hygiene, quality supervision, agriculture, 

trade and so on; trade markets, supermarkets, wholesale markets for agricultural products, hotels, 

large dining rooms, bases for agricultural production and so on.

 

Characteristics

•Without any calculating, showing the inspection result automatically.

•Supporting storage of 500 records and transmission of USB.

•Through testing of enzyme inhibition method.

•Storing supplementary reagent in the refrigerator of 4°C-6°C, low cost.

•Using half permanent light source, long life.

•Exquisite design, easy to carry.

 

Technical parameter

Power: 3.7V lithium battery inside, 5V power outside

Storage: 500 records (please upload the figure to epistatic machine when the saved figure will

reach the upper limit, or the figure will lose, test after deleting the figure in host)

Display: LCD screen 128×64

Transmittance accuracy: ±3%

Transmittance repeatability: ±2%

Linear error: ±5%

Button: Panasonic soft button

Others: display of time and electricity, set of automatically power off

Mode of display: LCD

PC interface: USB

Working environment: temperature: 0~40°C, humidity: 35~85%