BOD-573
BOD Analyzer
BOD-573 BOD Analyzer
BOD is a kind of environmental pollution index, which is mainly used to determine the pollution status of organic matter in water body; it can also be used as a basis to evaluate the operation efficiency of sewage treatment plant, so as to find the righttreatment method.
The method is based on pressure measurement in a closed system: microorganisms in the sample consume the oxygen and form CO2. This is absorbed by NaOH, creating a vacuum, which can be read directly as a measured value in mg/l BOD.
GENERAL FEATURES
● Measuring principle: differential pressure (manometric)
● The BOD value of the current test can be read at all times during or after the period of 5 days.
● BOD measurement ranges of up to 4,000 mg/l can be measured using different sample volumes.
INCLUDES
● BOD Analyzer base Unit
● BOD test terminal
● 6 BOD culture bottles(580ml)
● Stirrer
● Chemical cup
● Culture bottle holder
TECHNICAL PARAMETERS
Model |
BOD-573 |
|
Parameters |
BOD5 |
|
BOD |
BOD Range |
0-4000mg/L |
Accuracy |
Qualified for BOD5 accuracy test (Glucose-Glutamic acid standard solution BOD 5 =180-230mg/L)) |
|
Measurement |
Measurement days |
1-30 days |
Number of sample measurements |
≤ 6 |
|
Volume of container |
580ml |
|
Working Temperature |
20 ± 1 ℃ |
|
Recording interval |
6 minutes to 3 hours / time |
|
Data management |
Data Storage |
10-year BOD 5 result value |
Outputs |
Printing mode |
wireless transmission |
Display options |
Backlight |
Yes |
Auto Shutdown |
1~60 min, off |
|
Date and Time |
Yes |
|
General |
Power |
AC Adapter, 100-240V AC input |
Dimensions |
272mm × 185mm × 75mm |
|
Weight |
2400g |
|
Dimensions (Carton) |
490mm × 410mm × 220mm |
|
Weight (Gross) |
8 kg |