KD-3358
Semi-automated Rotary Microtome
Features:
This product adopts a streamline design, and possesses advanced features with stable performance and easy operation;
Specimen feed drive is controlled by an intelligent, high precision control system, enabling semi-ultrathin slicing, starting from a thickness of 0.25μm;
Specimen feeding drive utilizes an intelligent control system, it has high precision on cutting
Advanced drive systems; multiple functions including sectioning, trimming, fast forward, fast backward, conversion etc.;
Adopts imported cross-roller guide rails and a screw motion mechanism to ensure precision, thus achieving superior sectioning performance;
LCD screen shows section and trimming thickness, number of sections and trims;
Sectioning and trimming are completed through the control system;
equipped with a safety alert system, drive-overload protection mechanism and an automatic standby protection mechanism;
Easy and fast switching between different specimen clamps (two options: paraffin block clamp and cassette clamp);
Hand-wheel balance is precisely adjusted and can be locked at any position to ensure the safety and convenience of operation;
Blade holder can be laterally moved to avoid moving the blade with direct contact, thereby utilizing the entire length of blade (three different segments);
The red bar on the blade holder covers the whole length of blade to protect the user and the push bar enables the easy changing of the blade;
Hand-wheel balance is precisely adjusted and can be locked at any position to ensure the safety and convenience of sectioning;
Large-volume waste tray is easy to remove, and items such as deposable blades can be stored on the top of the housing.
Technical Specifications:
Section Thickness Setting Range: 0 - 60μm
Trimming Thickness Setting Range: 0 - 100μm
Section thickness adjustment:
0 - 1μm increment 0.25μm
1 - 10μm increment 1μm
10 - 20μm increment 2μm
20 - 60μm increment 5μm
- Total Horizontal Specimen Feed:26 mm
- Vertical Specimen Stroke: 52 mm
- Specimen Clamp Rotation: any angle within 360 degrees
- Specimen Orientation: XY – 8°
- Automatic return of specimen
- Minimum Setting of Sectioning Thickness: 0.25μm
- Maximum Specimen Size:50×50mm
- Precision Error: ±1%
- Working Voltage: AC 220V±10% 50Hz (standard model); AC110V±10% 60Hz
- Power: 100W
- Dimensions: 580×410×305mm(W×D×H)
- Net weight: 29kg