WS-150YDB/WS-200YDB/WS-280YDB
Horizontal cylindrical pressure steam sterilizer (microcomputer control)
Use
The horizontal autoclave, which adopte the way of gravity exchange to release the cold air from the chamber more completely to ensure the reliable sterilization. The control system automatically adjust the steam inlet and outlet according to chamber temperature during sterilizing. This unit is an ideal equipment for clinics, scientific research institutes and other organizations to sterilize surgical instruments, fabric glassed, and the culture media etc.
Characteristics:
1.Automatially controlled the sterilizing process, easy to operate.
2.With drying function, suitable dressing drying.
3.With over-temperature, over-pressure auto-protect device.
4.The door opening mechanism cannot be operated until the pressure in the chamberis reduced to 0.027MPa.It can’t be start on if the doesn’t closed well.
5.The safety value will be open automatically when the inner pressure over 0.24MPa,and the steam be exhausting to the water tank.
6.Automatically cut off the power if the water-lack happened, and alarm at the meanwhile.
7.The chamber of the sterilizer is made of stainless steel.
8.With printer.
Technical data
Technical data | WS-150YDB | WS-200YDB | WS-280YDB |
Sterilizing chamber | 150L | 200L | 280L |
volume | φ440×1000mm | φ515×1000mm | φ600×1000mm |
working pressure | 0.22 MPa | ||
working temperature | 134℃ | ||
Adjustment of temperature | 40℃ - 134℃ | ||
Time of sterilization | 0-99 min | ||
Time of drying | 0-99 min | ||
Heat average | ≤± 2℃ | ||
Power | 9KW/380V 50Hz | 9KW/380V 50Hz | 12KW/380V 50Hz |
Dimension | 1400×650×1820 | 1400×750×1700 | 1520×910×1900 |
Transport dimension | 1560×860×1820 | 1560×910×1880 | 1680×1060×2050 |
G.W/N.W | 430/340 KG | 436/350 KG | 570/462 KG |