PH500T
Benchtop PH Meter
GENERAL FEATURES
● Color High contrast LCD touchscreen, 7.0 inches.
● Intelligent operation system provides features including user management, calibration management, method management, data management and etc.
● Multi-reading feature allows auto-read, timed-read and continuous-read.
● Automatic/Manual temperature compensation ensures accurate results.
● Data storage 1000 sets for each parameter (GLP-compliant).
● Data analysis feature helps user review, compare and recalculate results.
● Support for USB or RS-232 communication.
● Reset feature automatically resumes all settings back to factory default options.
● IP54 waterproof.
● Automation measurement.
pH
● 1-6 points calibration with standard recognition.
● Selectable pH buffer groups, including DIN, NIST, USA, MERK, JIS, GB.
● Support for pH standard group customization.
● Automatic electrode diagnosis with pH slope and offset display
Conductivity
● 1-5 points calibration with standard recognition.
● Settable parameters, including cell constant, reference temperature (5/10/15/18/20/ 25°C), temperature compensation coefficient and TDS factor.
● Temperature compensation type (none, linear, pure water).
Ion
● 1-6 points calibration
● Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, ppm, etc.
● Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.
● Over 10 methods are built-in, including F- , Cl- , Br- , I- , NO3- , BF4- , NH4+ , K+ , Na+ , Ca2+, Cu2+, Pb2+, Ag+ and etc.; user-defined method is supported.
DO
● Support for air-saturated water or zero oxygen calibration.
● Auto barometric pressure compensation.
● Manual Salinity Factor Correction.
● Selectable pressure unit, including kPa, mbar, Torr, Atm.
Specifications
Model |
PH500T |
● |
|
Parameters (up to 12) |
pH Temp.(mvORP) |
● |
|
pH |
Range |
-2.000 to 20.000 pH |
● |
Resolution |
0.1,0.01,0.001pH |
● |
|
Accuracy |
±0.002 pH |
● |
|
Calibration Points |
Up to 6 |
● |
|
Standard Customization |
Yes |
● |
|
Calibration Reminder |
Yes |
● |
|
Standard Recognition |
NIST,GB,DIN,USA,MERK,JIS buffers |
● |
|
Slope Limit |
Yes |
● |
|
mV/ORP |
Range |
-2000.00 to 2000.00 mV |
● |
Resolution |
0.1,0.01 mv |
● |
|
Accuracy |
±0.1 mV or±0.03% ofreading whichever is greater |
● |
|
EH ORP Mode |
Yes |
● |
|
ORP Calibration Points |
1 custom point (relative mV) |
● |
|
pX |
Range |
-2.000 to 20.000 pX |
/ |
Resolution |
0.1,0.01,0.001pX |
/ |
|
Accuracy |
±0.002 pX |
/ |
|
Calibration Points |
Up to 6 |
/ |
|
ISE |
Range |
1E-9 to 9.999E9 |
/ |
Units |
molL,mmol/L, g/L, mg/L, ug/L,ppm |
/ |
|
Resolution |
Up to 4 significant digits |
/ |
|
Accuracy |
±0.3% |
/ |
|
Calibration Points |
Up to 6 |
/ |
|
Conductivity |
Range |
0.000 μS/cm to 2000 mS/cm |
/ |
Resoluton |
0.001 μS/cm minimum;changed with range |
/ |
|
Accuracy |
±0.5% FS |
/ |
|
Reference Temperature |
5,10,15,18,20,25C |
/ |
|
Calibration Points |
Up to 5 |
/ |
|
Calibration Reminder |
Yes |
/ |
|
Standard Recognition |
10uS/cm, 84μS/cm, 500uS/cm,1413μS/cm;12.88mS/cm |
/ |
|
Resistivity |
Range |
5.00Q cm~ 100.00MΩ-cm |
/ |
Resolution |
0.01Q-cm minimum |
/ |
|
Accuracy |
±0.5%FS |
/ |
|
TDs |
Range |
0.000~1000ppt |
/ |
Resolution |
0.001ppm minimum; changed with range |
/ |
|
Accuracy |
0.50% |
/ |
|
Salinity |
Type |
NaCI% |
/ |
Range |
(0.00~8.00)% |
/ |
|
Resolution |
0.01% |
/ |
|
Accuracy |
±0.1% |
/ |
|
DissolvedOxygenConcentration |
Sensor Type |
Polarographic |
/ |
Range |
0.00 to 99.99 mg/L |
/ |
|
Resolution |
0.01mg/L |
/ |
|
Accuracy |
±0.3mg/L(0.00 to 20.00 mg/L)±10%(20.00 to 99.99 mg/L) |
/ |
|
Calibration Reminder |
Yes |
/ |
|
Calibration Points |
Air-saturated water or zero point |
/ |
|
Barometric Compensation |
Yes |
/ |
|
Manual Salinity Factor Correction |
Yes |
/ |
|
% Saturation |
Range |
(0.0 to 600)% |
/ |
Resolution |
0.10% |
/ |
|
Accuracy |
±10.0% |
/ |
|
Temperature |
Range |
-10 to 135°C.14 to 275F |
● |
Unit |
C,F |
● |
|
Resolution |
0.1 |
● |
|
Accuracy |
±0.1 |
● |
|
Measurement |
Reading Mode |
AutoRead( Fast, Medium, Slow), Timed, Continuous |
● |
Reading Prompts |
Reading, Stable, Locked |
● |
|
Temp. Compensation |
ATC. MTC |
● |
|
DataManagement |
Data Storage |
1000 Groups |
● |
GLP Features |
Yes |
● |
|
Log Management |
Yes |
● |
|
Inputs |
pH Electrode |
BNC(Q9) |
● |
Temp./DO Probe |
4-pin aviation connector |
● |
|
Temp.EC.Probe |
5-pin aviation connector |
/ |
|
Outputs |
USB |
USB 2.0 flash memory device,PC |
● |
RS 232 |
printer、scanner、autosampler |
● |
|
DisplayOptions |
Backlight |
Yes |
● |
Auto Shut-down |
1~60 min,off |
● |
|
IP Rating |
IP54 |
● |
|
Date and Time |
Yes |
● |
|
General |
Power |
AC Adapter, 100-240 VAC input,DC24V output |
● |
Dimensions |
220x 195x68 mm |
● |
|
Weight |
950g(2.09 Ib) |
● |
± |
± |
± |